USER MANUAL8500 SERIES DC ELECTRONIC LOADSModels 8500, 8502, 8510, 8512, 8514, 8518, 8520, 8522, 8524 & 8526
Shift + Menu Enter the instrument's menu system.Shift + Short Turn short circuit on or off.Shift + Tran Start or stop transient condition.Shift +
not covered by warranty.4Line voltage selection switch (110 VAC or 220 VAC)Display annunciatorsAnnunciator MeaningOFF The load is off.UNREG The input
Lvl Menu item Function2 INITIAL CONFIG Return instrument to factory default settings.2 POWER-ON RECALL Power on state of instrument.3 ON Remembers sta
Lvl Menu item Function2 BAUDRATE SET Set the serial bus connection speed.348003 96003 192003 384002 COMM. PARITY SET Serial connection parity.3NONE3 E
Lvl Menu item Function4 REPEAT List is repeatedly executed after the trigger is received.3 VOLTAGE LIST4 ONCE List is executed once per trigger.4 REPE
SpecificationsModels 8500 & 8502 (300 W)Parameter 8500 8502Input rating Voltage 0 to 120 V 0 to 500 V Current 1 mA to 30 A 1 mA to 15 A
Models 8520/8522/8524/8526 (2400W & 5000W)Parameter 8520 8522 8524 8526Input rating Voltage 0 – 120 V 0 – 500 V 0 – 60 V 0 – 500 VCurrent 0 – 240
Parameter SpecificationHumidity ≤ 95% relative humidity, non-condensingAltitude ≤ 2000 mAC Line voltage220 AV±10%,47~63 Hz110 AV±10%,47~63 HzOperating
The curved portion is where the dissipated power is at the rated power of the instrument (and is actually a hyperbolic shape). When you use the menu
Slew rateThe slew rate for each DC load varies from model to model. The variations are also dependent on the different regions measured for each indi
SafetyThe following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precau
As a reference, below is a table of slew rates for their respective models:Model Slew rate85000.5A / S85020.5A / S85101A/ S85120.5A / S85141A/ S8
Glossary△ Up arrow key. Used to scroll through the menu or cause the temporary display of the alternate standard display.▽ Down arrow key. Used to s
R-set Configure the instrument for constant resistance mode.Recall Recall instrument state from non-volatile memory.Remote sensing Allows the instrume
InstallationInspectionItems you should have receivedWhen you open the box containing the instrument, you should find the following items:1. The DC Loa
Models 8510, 8512, 8514 & 8518Models 8520, 85228500 DC Load Series Version: November 4, 2008 Page 24 of 76
Models 8524 & 8526Unit (mm)8500 DC Load Series Version: November 4, 2008 Page 25 of 76
Model number Dimensions in mm Mass in kg8500 215W×88H×355D 5.28502 215W×88H×355D 5.28510 429W×88H×355D 148512 429W×88H×355D 148514 429W×88H×355D 14851
First turn-on checkoutCAUTIONEnsure that the line voltage selector switch on the back panel is set to match your line voltage. Failure to do so could
watt.Press the R-set button (you may have to press it twice). Set the resistance to 100 Ω. Press the On/Off key. Verify that the current is approx
Power-on stateThe default power-on instrument state is to not remember the instrument mode settings before the last power-down. If you wish to have t
Table of ContentsSafety...
Key DisplayI-set If you press this key after powering up and there is no currently-stored constant current value, you'll be prompted for the desi
Key Displaypower level. If there was a power value already stored, it will be momentarily displayed and the instrument will be in CW mode. If you wi
Key DisplayEnter :TIMER STATEEnter Use arrow keys to set to :ON.▽ :TIMER SETEnter TIMER=XXXXXS Enter desired time interval. Valid values are 1 to 60
the battery) in ampere*hours (A*hrs) of the battery is calculated and displayed. To run a battery test, follow these steps:Key presses DisplayI-set,
Keys DisplayShift + S-Tran LEVEL A= X.XXXA You are being prompted for the first current value. Press the 5 key for 5 A.Enter Enters the 5 A value, th
Keys Displaycurrent value. Press the 1 and 0 keys to enter 10 A.Enter Enters the 10 A value, then displays WIDTH B = X.XMS. You're being promp
Keys DisplayEnter Displays LEVEL B= X.XXXA. You are being prompted for the second current value. Press the 1 and 0 keys to enter 10 A.Enter Enters t
Current, A Duration, ms Between times3 1000 0 and 10 800 1 and 22 500 2 and 30 300 3 and 46 500 4 and 5This list has 5 steps, located at the transitio
modes, mode parameters, and durations. They are useful for executing a set of tests on a device, then displaying whether the tests passed or failed.
Keys Display2.5 Enter SHORT OFF We select short on and then press Enter. Enter▽ READBACK A We want to readback the current, so we press Enter.Enter
Triggering...39
Trigger Type Explanationfront panel keys.EXTERNAL An external trigger is a TTL high signal applied to the Trigger connection on the back panel. This
front panel using the I-set, V-set, P-set, or R-set keys.To set these protection values, use the following keystrokes:Keys DisplayShift + Menu :CONFIG
reversed polarity. If a reverse voltage condition is detected, the buzzer will sound and REVERSE VOLTAGE will be displayed on screen.Over Temperature
Example: A power supply is connected to the DC load with 72.5 cm of 20-gauge solid copper wires. The constant current i set to 5 A. The power supply
Remote operationCommunication cablesThe DC Load has a DB9 connector on the rear panel that allows remote communication.WARNINGDo not connect the DC Lo
This COM port can then be accessed as if it were a regular RS-232 port.The LEDs in the adapter will blink when information is being sent through the a
The checksum number is the arithmetic sum of each of the bytes modulo 256.Status packetsWhen you send a command that does not cause the DC Load to sen
print " ", s = "%02x" % ord(bytes[i]) if s == "00": s = chr(250)*2 print s,
Here are the printed results when the above script is ran:Set to remote command: aa ·· 20 01 ·· ·· ·· ·· ·· ·· ·· ·· ·· ·· ·· ·· ·· ·· ·
example programs. The Summary of commands section is a list of the commands, but without details. The Command details section explains how to use ea
NotationI-set Represents a key on the front panel. Note that some of these may be accessed in combination with the Shift key.OFF Annunciator, value,
Command Group Byte ActionList operations0x3A Select the list operation (CC/CV/CW/CR)0x3B Read the list operation (CC/CV/CW/CR)0x3C Set how lists repea
Command Group Byte Action0x5C Recall DC Load's settingsFunction0x5D Select FIXED/SHORT/TRAN/LIST/BATTERY function0x5E Get function type (FIXED/SH
Least significant two bytes, most significant byte Lower high byteMost significant two bytes, least significant byte Higher low byteMost significant t
Byte offsetMeaning4 Lower high byte of maximum voltage.5 Upper low byte of maximum voltage.6 Upper high byte of maximum voltage.7-24 ReservedExample:
Byte offsetMeaning5 Upper low byte of maximum current.6 Upper high byte of maximum current.7-24 Reserved0x26 Set the maximum power allowedByte offse
Byte offsetMeaning3 Mode:0 is CC1 is CV2 is CW3 is CR4-24 Reserved0x2A Set CC mode current Byte offsetMeaning3 Lower low byte of current. 1 represe
Byte offsetMeaning4 Lower high byte of voltage.5 Upper low byte of voltage.6 Upper high byte of voltage.7-24 Reserved0x2E Set CW mode powerByte offs
Byte offsetMeaning5 Upper low byte of resistance.6 Upper high byte of resistance.7-24 Reserved0x32 Set CC mode transient current and timing Byte off
Byte offsetMeaning0 is CONTINUOUS1 is PULSE2 is TOGGLED16-24 Reserved0x35 Read CV mode transient parametersByte offsetMeaning3 to 6 Value A of volta
Byte offsetMeaning15 Transient operation:0 is CONTINUOUS1 is PULSE2 is TOGGLED16-24 Reserved0x38 Set CR mode transient resistance and timingByte off
Quick referenceModel numbers covered by this documentThe B&K Precision DC Loads covered by this manual are:● 8500● 8502● 8510● 8512● 8514● 8518●
Byte offsetMeaning3 is constant resistance (CR)4-24 Reserved0x3B Read the list operation (CC/CV/CW/CR)Byte offsetMeaning3 List operation mode:0 is c
0x40 Set one of the step's current and time valuesByte offsetMeaning3 to 4 2 byte little-endian integer specifying which step number in the lis
0x45 Set one of the step's power and time valuesByte offsetMeaning3 to 4 2 byte little-endian integer specifying which step number in the list5
Byte offsetMeaning3 Partition scheme:1 means 1 file of 1000 list steps2 means 2 files of 500 list steps4 means 4 files of 250 list steps8 means 8 file
Byte offsetMeaning7-24 Reserved0x50 Set timer value of for LOAD ONByte offsetMeaning3 to 4 2 byte little-endian integer specifying the time in units
Byte offsetMeaning1 means to enable the Local key on the front panel4-24 Reserved0x56 Enable/disable remote sensingByte offsetMeaning3 0 means to di
Byte offsetMeaning3 Storage register, a number between 1 and 25 inclusive4-24 Reserved0x5C Recall DC Load's settingsByte offsetMeaning3 Storage
The operation state register's bit meanings are:Bit Meaning0 Calculate the new demarcation coefficient1 Waiting for a trigger signal2 Remote cont
Byte offsetMeaning3 Calibration protection state:Bit 0: value of 0 means the calibration state is not protectedBit 0: value of 1 means the calibrati
The stored calibration data will be used the next time the DC Load is power cycled.0x67 Set calibration informationByte offsetMeaning3 to 22 ASCII i
whose resistance doesn't change as a function of current or voltage).The DC Load can present dynamically changing loads to the DC source with mil
Byte offsetMeaning12-24 Reserved8500 DC Load Series Version: November 4, 2008 Page 70 of 76
Serial number and firmware versionTo find out the serial number and firmware version of the DC load, turn the instrument on. While the SYSTEM SELFTES
In case of troubleInstrument won't turn onIf the instrument won't turn on when the POWER switch is pressed in, please ensure the power cord
Enter again. Now the password is cleared. Press Esc to exit the menu To unlock the other DC load models, use the following unlock keys:Model number
Appendix: Service and warranty information SERVICE INFORMATION Warranty Service: Please go to the service and support section on our website www.bkp
IndexAlphabetical IndexBattery...21Battery test...
22820 Savi Ranch Parkway Yorba Linda CA, 92831Printed in China8500 DC Load Series Version: November 4, 2008 Page 76 of 76
6123 4 5Button Function(s)116-character display shows voltage and current measurements.2Rotary knob. Turn to adjust a setting value. Press in to to
Standard displayThe standard display for the instrument is the voltage and current at the instrument's terminals. You may press the △ or ▽ keys
Kommentare zu diesen Handbüchern