Byte
offset
Meaning
3 Partition scheme:
1 means 1 file of 1000 list steps
2 means 2 files of 500 list steps
4 means 4 files of 250 list steps
8 means 8 files of 120 list steps
4-24 Reserved
0x4B Read the memory partitioning for storing list steps
Byte
offset
Meaning
3 Partition scheme:
1 means 1 file of 1000 list steps
2 means 2 files of 500 list steps
4 means 4 files of 250 list steps
8 means 8 files of 120 list steps
4-24 Reserved
0x4C Save the list file
Byte
offset
Meaning
3 Storage location, a one byte integer from 1 to 8. This number must be
consistent with the number of list files allowed as set by the 0x4A command.
4-24 Reserved
0x4D Recall the list file
Byte
offset
Meaning
3 Storage location, a one byte integer from 1 to 8.
4-24 Reserved
0x4E Set minimum voltage in battery testing
Byte
offset
Meaning
3 to 6 4 byte little-endian integer specifying the minimum voltage in units of 1 mV
7-24 Reserved
0x4F Read minimum voltage in battery testing
Byte
offset
Meaning
3 to 6 4 byte little-endian integer specifying the minimum voltage in units of 1 mV
8500 DC Load Series Version: November 4, 2008 Page 63 of 76